首頁 >> 供求商機
進口2950型雙波長掃描紫外可見光度計
model 2950 double beam pc scanning spectrophotometer
model 2950 is a high performance uv-vis double beam automatic scanning spectrophotometer. it is a two (2) cell spectrophotometer with a fixed bandwidth of 2.0 nm. model uvd-2950 spectrophotometer offers high performance, ease of use and reliability, which can be used in various applications. spectrophotometer model 2950 can be used extensively for qualitative and quantitative analysis in such fields as pharmaceutical inspection, clinical analysis, petro-chemistry laboratories, chemistry and biochemistry laboratories, dna/rna analysis as well as in quality control departments, i.e. environmental control, water management, food processing, and agriculture.model 2950 utilizes a new optical system design and is microcomputer controlled. with its focused-beam design, the system provides optimal and reproducible results for small samples. the sample beam and the reference beam are provided within the same sampling space which in turn facilitates wider and longer scan of data providing a more detailed view of the results in an easy to use environment. this instrument has excellent baseline stability and high resolution and permits scanning, quantitative analysis, kinetic spectrophotometric analysis and dna/rna analysis through pc control. this product is capable of processing data, from analytical and spectrum testing. model 2950 has a large lcd screen which displays the menu screen and of course makes the device user friendly. additionally, this instrument has a powerful built-in software which permits the apparatus to be linked to a computer and a printer to display the photometric and spectral data on the pc monitor.our company. is certified by iso 9001-2000, has ce conformity and is fda licensed. model 2950 with fixed bandwidth of 2 nm is a high-performance, reliable, and exceptional value instrument which is the hallmark of our uv-vis spectrophotometers.
2950型雙波長掃描紫外可見光度計
●features
baseline stability: the double beam monitoring ratio system enhances baseline stability.
excellent resolution: the big-caliber light path enhances the instrument’s energy, reduces its noise and raises its resolution performance
2 cell holder: model 2950 has 2 cell holder for reference (standard) and sample.
user-friendly light source: the socket deuterium lamps and tungsten lamps facilitate light source replacement, simplify maintenance and reduce operation error.
convenient display: the large backlit lcd screen displays both photometric values and spectral curves.
full use of computer technology: being computer controlled with rs-232 interface and working on the windows platform with the uv/win application software.
the key components adopt all from the world famous manufacturer, such as deuterium lamp, silicon photodiode and holographic grating, which ensures the stabilization and credibility of the instrument for extended life.
computer system is optional (not included).
●software specifications
monoprocessor built-in application:
photometric measurement: measuring transmittance or absorbance at the current wavelength together with k factor calculations.
spectrum scan: carrying out scanning of transmittance or absorbance on the se-lected wavelength range together with peak-pick module.
quantitative determination: regression of standard curves and direct determination concentration of samples.
pc windows application software (rs-232 interface) to link spectro to computer and printer:
photometric measurement: measuring the photometric values at 1-10 wavelengths together with mathematical calculations according to entered quotations.
spectrum scan: producing wavelenght scans within the operating parameters on samples together with powerful data handling facilities.
quantitative determination: determination of unknown concentration with methods of 1-3 wavelength quantitation together with fitting of calibration curve of 1st ~ 4th order.
kinetics: recording curves of changing photometric values of samples against time course at the se-lected wavelengths together with powerful data handling facilities.
output: with the windows clipboard, the measured data and graphics can be copied to other applications software for reports.
●included accessories
2 auto cell holder
4 optical glass cells 10mm.
2 quartz cells 10mm.
1 dust cover
1 instruction manual
1 power cable
1 pc cable
1 software cd for windows 98/2000/xp
1 software operation manual
1 spare tungsten halogen lamp
1 block light cell
1 extra fuse
optional: peltier kinetic test system
optional: sipper flow through system
●optional accessories
set of 2 performance testing filters (1 "e filter" for photometric accuracy test and 1 didymium filter for wavelength accuracy test)
peltier constant temperature system (15 - 55 ºc specify: spectro uv-vis double)
sipper flow through system (peristaltic pump and flowcell (specify: spectro uv-vis double)
multi-purpose cell holder 20-30-40-50 path length
●technical specifications
1)wavelength range: 190 to 1100 nm
spectral bandwidth: 2nm
resolution: 2nm
stray light: 0.2%t (220 nm and 340 nm)
wavelength accuracy: 0.5 nm (with automatic wavelength correction).
wavelength repeatability: 0.2nm
2)photometric system: the double-beam monitoring ratio system
photometric method: transmittance, absorbance, energy and concentration
photometric range: 0.3~3.0 abs (0~200%tt)
photometric accuracy: 0.002abs (0~0.5abs) , 0.004abs (0.5~1.0abs), ±0.3%t (0-*t)
photom. reproducibility: 0.001abs (0~0.5 abs), 0.002abs (0.5~1.0abs), 0.15%t (0~*t),
photometric display: -9999 ~ 9999
photometric noise: <±0.001abs (500nm, 0abs, 2nm bandwidth)
scanning speed: 1400nm/min
baseline flatness: 0.002abs (190 nm. ~1100 nm.)
baseline stability: 0.002abs/h (500 nm., after preheating).
slew rate of wavelength: 3600nm/min
3) dna/rnd measurement:
results printout: printing of measured data by using any printer with paralell port connection available.
4)mainframe: compact and standalone spectrophotometer mainframe
light source: socket deuterium lamp and socket tungsten halogen lamp
detector: double beam
sample chamber: 2 cell holder
display: liquid crystal display (lcd 320 ià240 dot matrix)
keypad: touch soft keys.
pc interface: pc interface: rs-232
size: 22x16x10"
weight: 55 lb