產(chǎn)品簡(jiǎn)介

靈活的PH微電極
Calibration (In-vitro Calibration)
The Flexible pH Electrode along with a standard laboratory calomel or silver-silver chloride reference
electrode with liquid junction,

詳細(xì)介紹

靈活的PH微電極

The Flexible pH Electrode along with a standard laboratory calomel or silver-silver chloride reference
electrode with liquid junction, can be used as the electrode pair. The pair is calibrated in at least two pH
buffer solutions which are close to the pH values expected. Follow the procedure recommended by the
manufacturer of your pH meter for calibrating your pH meter with our MI-506/MI-508.
Optimum Response Time: Optimum response time will be obtained after the probe has been exercised in
two buffer solutions. Place a pH 4.01 buffer or equivalent in a beaker and a pH 6.86 buffer or equivalent
in a second beaker.
Hold the MI-506/MI-508 and reference electrode together and touch the pH 4.01 buffer allowing 15-20
seconds for equilibration. Rinse the two electrodes with distilled water and then touch the pH 6.86 buffer
in the same manner. Do this several times.
A skin reference (i.e. our A-30) is applied to the body of the patient and the pH electrode is placed into a
test tube containing a pH 7 buffer solution. The patient touches either the buffer solution or a buffer
soaked gauze extending from the buffer solution. The pH 7 reading can now be calibrated on the meter.
A second pH electrode is now placed into this second test tube and the patient must now make contact
with this buffer as mentioned above. The pH electrode can also be calibrated on the technician before
measurements on a patient. However, this method of calibration can result in actual pH error up to 0.5
pH units.

靈活的PH微電極
Sterilization and DisinfectionThe electrode can be sterilized in ethylene oxide or chemically sterilized with gluteraldehyde solutions such
as Cidex 7.
Note: Probe cannot be autoclaved. The electrode can be disinfected with ethyl
alcohol or zephirin chloride.
Accepted methods of sterilization should be followed prior to usage.
Handling
Avoid excessive bending or sharp angle bends just behind the rigid (bulb) tip.
Be careful not to apply pressure against the inner glass capillary tube.

靈活的PH微電極

Cleaning
When using the microelectrode in solutions containing protein, it should be soaked in an enzyme cleaning
solution such as Terg-a-zyme (Alconox, Inc.) or a chromic sulfuric glass cleaning solution can be used to
remove the protein build up.
Note: When cleaning, never pinch the electrode between fingers and pull
through for drying --- always pat dry.
After cleaning, rinse thoroughly with distilled water.
Storage
Always clean the microelectrode before storing:
Long-term (over 2 weeks): over 2 weeks): Return the electrode to its original container and prepare it
in the same condition in which you received it.
Short-term: The probe can be left in an acid pH buffer solution (pH 4.01).