Conventional transducers measure the strain in a compliant member and based on their stiffness, determine the torque resp. normal force (Fig. 1a). This method has several inherent disadvantages. Since the compliant member is strained, the strain has to be accounted for in determining actual sample strain. In dynamic the inertia of the moving mass of the transducer, affects the measured results at high frequences. Further, viscoelastic materials which develop normal forces, drive the cone and plate apart. This effect generates secondary flow fields and the result is an inaccurate transient normal force measurement. To overcome these problems of conventional transducer, all developments have been made towards stiffer transducers. The stiffness however is directly related to the resolution of the strain measurement of the compliant member i.e. the torque resp. force measurement. This implies that the stress resolution decreases with increasing stiffness.
提供商 |
TA儀器 | 下載次數(shù) |
300次 |
資料大小 |
190.4KB | 資料類(lèi)型 |
PDF 文件 |
資料圖片 |
-- | 瀏覽次數(shù) |
1102次 |
(空格分隔,最多3個(gè),單個(gè)標(biāo)簽最多10個(gè)字符)
立即詢(xún)價(jià)
您提交后,專(zhuān)屬客服將第一時(shí)間為您服務(wù)