您好, 歡迎來到化工儀器網(wǎng)

| 注冊(cè)| 產(chǎn)品展廳| 收藏該商鋪

18217040487

Download

首頁   >>   資料下載   >>   力平衡傳感技術(shù)

TA儀器

立即詢價(jià)

您提交后,專屬客服將第一時(shí)間為您服務(wù)

力平衡傳感技術(shù)

閱讀:1182      發(fā)布時(shí)間:2016-06-29
分享:

Conventional transducers measure the strain in a compliant member and based on their stiffness, determine the torque resp. normal force (Fig. 1a). This method has several inherent disadvantages. Since the compliant member is strained, the strain has to be accounted for in determining actual sample strain. In dynamic the inertia of the moving mass of the transducer, affects the measured results at high frequences. Further, viscoelastic materials which develop normal forces, drive the cone and plate apart. This effect generates secondary flow fields and the result is an inaccurate transient normal force measurement. To overcome these problems of conventional transducer, all developments have been made towards stiffer transducers. The stiffness however is directly related to the resolution of the strain measurement of the compliant member i.e. the torque resp. force measurement. This implies that the stress resolution decreases with increasing stiffness.

提供商

TA儀器

下載次數(shù)

300次

資料大小

190.4KB

資料類型

PDF 文件

資料圖片

--

瀏覽次數(shù)

1182次

產(chǎn)品展示

會(huì)員登錄

請(qǐng)輸入賬號(hào)

請(qǐng)輸入密碼

=

請(qǐng)輸驗(yàn)證碼

收藏該商鋪

標(biāo)簽:
保存成功

(空格分隔,最多3個(gè),單個(gè)標(biāo)簽最多10個(gè)字符)

常用:

提示

您的留言已提交成功!我們將在第一時(shí)間回復(fù)您~
在線留言