1. PERFORMANCE
1) Measuring range :1-60 ppm 0.5-1 ppm
Number of pump strokes 1(100ml) 2(200ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :0.2 ppm(200mr)
4) Shelf life :1 year
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Red
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Carbon tetrachloride is decomposed and Phosgene is produced. By reacting with this Phosgene,
4- (p-Nitrobenzyl)-phridine and Benzylaniline, dyestuff is produced.
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
DETECTOR TUBE
(white) (Olive)
PRETREAT TUBE
1
5
10
20
30
40
50
60
100ml
ppm
147S CARBON TETRACHLORIDE
Substance Interference ppm Coexistence
Phosgene Similar stain is produced. 2 Higher reading are given.
Halogenes FIG.1 〃
Halogenated hydrocarbonFsIG.2 〃
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
60 40 50 60 78 -
50 34 42 50 62 76
40 27 34 40 46 57
30 21 25 30 33 38
20 15 17 20 21 22
10 9 10 10 10 10
5 5 5 5 5 5
1 1 1 1 1 1
(NOTE)
When the concentration is below 1 ppm, 2 pump strokes can be used to determine the lower concentration.
Following formula is available for the actual concentration.
Actual concentration=1/2×Temperature corrected value
6.CAUTION
When high concentration gas(over the full scale)is sampled, the indicator will be discoloured from pale yellow
to dark blue. If extremely high concentration gas is sampled, the detecting reagent may be faded and it is not
possible to distinguish with original colour.
0
40
20
Chlorine〔ppm〕?
FIG.1 Influence of Chlorine
Carbon tetrachloride
indicating value 〔ppm〕
10 20 0
40
20
Trichloroethylene〔ppm〕?
FIG.2 Influence of Trichloroethylene
請(qǐng)輸入賬號(hào)
請(qǐng)輸入密碼
請(qǐng)輸驗(yàn)證碼
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購(gòu)買(mǎi)風(fēng)險(xiǎn),建議您在購(gòu)買(mǎi)產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。