當前位置:上海境宣實業(yè)有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 194S1,3-二氯丙烷氣體檢測管194S 1,3-DICHLOROPROPANE 10-500ppm
參 考 價 | 面議 |
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
1,3-二氯丙烷氣體檢測管194S 1,3-DICHLOROPROPANE 10-500ppm
Tube No.
194S
1. PERFORMANCE
1) Measuring range :10-500 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :1 ppm
4) Shelf life :1 year (Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, chlorine is produced. Yellowish Orthoquinone is produced by reacting
between this Chlorine and o-Toulidine.
CH2CICH2CH2CI+I2O4+CrO3+H2SO4→CI2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
??????????
GAS To pump
DETECTOR TUBE
(White)
PRETREAT TUBE
500
400
300
200
100
50
30
10
ppm
100ml
194S 1,3-DICHLOROPROPANE
Substance Interference Coexistence
Alcohols The accuracy of readings is not affected. The accuracy of readings is not
affected.
Esters 〃〃
Aromatic hydrocarbons 〃〃
Aliphatic hydrocarbonFsIG.1 〃〃
Halogenated hydrocarFbIoGn.s2 Similar stain is produced. Higher readings are given.
0
300
400
500
200
100
Trichloroethylene〔ppm〕?
FIG. 2 Influence of Trichloroethylene